Storage of electrochemical gas sensors:
1, The storage of the sensor shall not exceed 6 months and shall be stored in a sealed container with clean air between 0 and 20 degrees Celsius.
2, The sensor can not be stored in the environment with liquid vapor, all the sensors can not be applied in the environment containing organic vapor, otherwise it will destroy the performance of the sensor.
3, The sensor should not be stored and used in an environment with concentrated liquid, because in such an environment the liquid will block the pores, limit the flow of gas into the sensor and weaken the signal. If the sensor is affected by concentrated gas, dry the sensor with a piece of soft absorbent paper before re-operating. But under no circumstances should the sensor be dried at temperatures above 40 degrees Celsius.
4, Generally, sensors can be stored in the office environment. The temperature is not more than 30 degrees Celsius, and the humidity is 50%-90%.
5, Special storage sensor can use fresh-keeping box, followed by the refrigerator can also be used, drying box is not recommended. Placed in the fresh-keeping box and freezer sensor should also be sealed storage, effectively prevent because of multiple opening of the box, the environment of the water gas into the box to form condensation, blocking the sensor's porosity.